Indian Journal of Science and Technology
DOI: 10.17485/ijst/2016/v9i20/94683
Year: 2016, Volume: 9, Issue: 20, Pages: 1-5
Original Article
Dong-Hee Hong, Joon-Koo Choi*, Gha-Jung Kim and Hyeong-Gyun Kim
Department of Radiology Science, Far East University, Chungbuk, Korea; [email protected], [email protected], [email protected], [email protected]
*Author for correspondence
Joon-Koo Choi Department of Radiology Science, Far East University, Chungbuk, Korea; [email protected]
Background/Objectives: An objective of this study is to analyze the size of bore on signal to noise ratio in magnetic resonance imaging. Methods/Statistical Analysis: During this process, the magnetic field also shows irregularity due to the environmental factors. Therefore in this study, magnetic field non-uniformity according to the bore size of MRI will be examined as well as the influence on the image through SNR measurement. Findings: Avanto with bigger bore size has shown higher SNR value, and especially, it has shown 6 times or greater difference in TSE(Turbo Spin Echo) technique. Improvements: It would be imperative to conduct a test with a small bore size and a uniform magnet when the MRI test failure rate is high due to claustrophobia.
Keywords: Bore size, Claustrophobia, Magnetic field, Magnetic Resonance Image (MRI), Pulse sequence, Signal to Noise Ratio (SNR)